It is used to test the location of rebar in concrete, the thickness of the protective layer, and the diameter, orientation, and distribution of rebar by using an electromagnetic induction method. It also can be used to test the distribution of magnetic and conductor such as cables and metal pipes in non-magnetic medium and non-conducting medium.
Number
Item
SPEC
1
Bar diameter application range(mm)
Φ6 – Φ50
2
maximum value permissible error of Bar
diameter
≤±1
3
Protective layer thickness measuring
range
The first measuring range (mm)
6 - 90
The second measuring range (mm)
7-180
4
The maximum error of reinforced
protective layer thickness test value
The first measuring range (mm)
The second measuring range (mm)
6 - 59
7 - 79
≤±1
60 - 69
80 - 119
≤±2
70 -90
120 - 180
≤±4
5
Work environment condition
Work temperature ℃
-10 - +40
relative humidity RH
< 90%
Features
Quick, easy scanning of large areas
Individual scans over lengths of up to 30 m
Data transfer from scanner by infrared link for monitor viewing
Cordless scanner for maximum freedom of movement
Immediate high resolution image for clear Picture of the reinforcement